Reduced field-of-view diffusion imaging of the human spinal cord: comparison with conventional single-shot
G Zaharchuk1, E U Saritas, J B Andre
1Department of Radiology, Stanford University, California, USA. gregz@stanford.edu
Background And Purpose:
DWI of the spinal cord is challenging because of its small size and artifacts associated with the most commonly used clinical imaging method, SS-EPI. We evaluated the performance of rFOV spinal cord DWI and compared it with the routine fFOV SS-EPI in a clinical population.
Materials And Methods:
Thirty-six clinical patients underwent 1.5T MR imaging examination that included rFOV SS-EPI DWI of the cervical spinal cord as well as 2 comparison diffusion sequences: fFOV SS-EPI DWI normalized for either image readout time (low-resolution fFOV) or spatial resolution (high-resolution fFOV). ADC maps were created and compared between the methods by using single-factor analysis of variance. Two neuroradiologists blinded to sequence type rated the 3 DWI methods, based on susceptibility artifacts, perceived spatial resolution, signal intensity-to-noise ratio, anatomic detail, and clinical utility.
Results:
ADC values for the rFOV and both fFOV sequences were not statistically different (rFOV: 1.01 ± 0.18 × 10(-3) mm(2)/s; low-resolution fFOV: 1.12 ± 0.22 × 10(-3) mm(2)/s; high-resolution fFOV: 1.10 ± 0.21 × 10(-3) mm(2)/s; F = 2.747, P > .05). The neuroradiologist reviewers rated the rFOV diffusion images superior in terms of all assessed measures (P < 0.0001). Particular improvements were noted in patients with metal hardware, degenerative disease, or both.
Conclusions:
rFOV DWI of the spinal cord overcomes many of the problems associated with conventional fFOV SS-EPI and is feasible in a clinical population. From a clinical standpoint, images were deemed superior to those created by using standard fFOV methods.