Evaluation of Respiratory System Mechanics in Mice using the Forced Oscillation Technique
Published on: May 15, 2013
Respiratory mechanics measured by forced oscillations during mechanical ventilation through a tracheal tube
Alexander-Wigbert Scholz1, Norbert Weiler, Matthias David
1Department of Anaesthesiology, University Medical Centre of the Johannes Gutenberg-University, 55131 Mainz, Germany. ascholz@uni-mainz.de
Abstract:
The forced oscillation technique (FOT) allows the measurement of respiratory mechanics in the intensive care setting. The aim of this study was to compare the FOT with a reference method during mechanical ventilation through a tracheal tube. The respiratory impedance spectra were measured by FOT in nine anaesthetized pigs, and resistance and compliance were estimated on the basis of a linear resistance-compliance inertance model. In comparison, resistance and compliance were quantified by the multiple linear regression analysis (LSF) of conventional ventilator waveforms to the equation of motion. The resistance of the sample was found to range from 6 to 21 cmH(2)O s l(-1) and the compliance from 12 to 32 ml cmH(2)O(-1). A Bland-Altman analysis of the resistance resulted in a sufficient agreement (bias -0.4 cmH(2)O s l(-1); standard deviation of differences 1.4 cmH(2)O s l(-1); correlation coefficient 0.93) and test-retest reliability (coefficient of variation of repeated measurements: FOT 2.1%; LSF 1.9%). The compliance, however, was poor in agreement (bias -8 ml cmH(2)O(-1), standard deviation of differences 7 ml cmH(2)O(-1), correlation coefficient 0.74) and repeatability (coefficient of variation: FOT 23%; LSF 1.7%). In conclusion, FOT provides an alternative for monitoring resistance, but not compliance, in tracheally intubated and ventilated subjects.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation I: Indication and Settings
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)