Related Experiment Video
Updated: Jun 3, 2026

08:43
Diagnosis of Musculus Gastrocnemius Tightness - Key Factors for the Clinical Examination
Published on: July 7, 2016
A new device for assessing ankle dorsiflexion motion: reliability and validity.
Jason Wilken1, Smita Rao, Miriam Estin
1Brooke Army Medical Center, Fort Sam Houston, TX, USA.
The Journal of Orthopaedic and Sports Physical Therapy
|April 5, 2011
Summary
The Iowa ankle range of motion (IAROM) device accurately measures ankle dorsiflexion range of motion and stiffness. This custom device offers a reliable and valid clinical tool for assessing ankle function.
Area of Science:
- Biomechanics
- Clinical Measurement
- Orthopedics
Background:
- Limited ankle dorsiflexion is linked to foot pain in various patient groups.
- Accurate assessment of ankle dorsiflexion is crucial in foot and ankle examinations.
- Traditional goniometric methods show poor intertester reliability, necessitating improved measurement tools.
Purpose of the Study:
- To evaluate the validity and reliability of the Iowa ankle range of motion (IAROM) device.
- To assess the IAROM device's capability in measuring ankle dorsiflexion range of motion (ROM) and stiffness.
- To establish the clinical utility of a novel, cost-effective device for ankle assessment.
Main Methods:
- The IAROM device's validity and intertester reliability were tested using applied dorsiflexion torque (10-25 Nm).
- Ankle dorsiflexion ROM and stiffness were measured with the knee extended and flexed.
- Convergent validity was assessed against an optoelectronic system, and intertester reliability was evaluated by two clinicians.
Main Results:
- The IAROM device demonstrated excellent validity for ankle dorsiflexion ROM (ICC: 0.95–0.98).
- Good to excellent intertester reliability was observed for ROM measurements (ICC: 0.90–0.95).
- Reliability for ankle joint dorsiflexion stiffness measurement was good (ICC: 0.71–0.85).
Conclusions:
- The IAROM device is a valid and reliable tool for measuring ankle dorsiflexion ROM.
- The device enables stiffness calculation, though its reliability for stiffness measurement is not optimal.