Related Experiment Video
Updated: Jun 3, 2026

Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Bio-separation, speciation and determination of chromium in water using partially pyrolyzed olive pomace sorbent
Amjad H El-Sheikh1, Malek M Abu Hilal, Jamal A Sweileh
1Department of Chemistry, Faculty of Science, Hashemite University, P.O. Box 150459, Zarqa 13115, Jordan. amjadelsheikh3@yahoo.com
Abstract:
Speciation of Cr(III)/Cr(VI) from water using olive pomace (OP) was improved by partial pyrolysis of OP. The sorbents were characterized by physicochemical techniques. Sorption of Cr(III) on raw and partially pyrolyzed OP sorbents followed Freundlich isotherm and second-order rate kinetics. OP pyrolyzed at 150°C (sorbent OP-150) exhibited maximum sorption capacity, favorability and the lowest sorption energy. Sorption was exothermic and spontaneous for the raw-OP and OP pyrolyzed at 100 or 150°C; but endothermic and non-spontaneous for OP pyrolyzed at 200, 250, 300 or 400°C. A speciation method of chromium was proposed, in which Cr(III) was selectively retained at pH 3 on sorbent OP-150; while total Cr was determined after reduction of Cr(VI). The method was selective with a detection limit for Cr(III) of 1.58 μg L(-1). The method was applied on natural and industrial waters (recoveries >97.7%, RSD's <9%) and on tobacco leaves certified reference material (INCT-PVTL-6).
More Related Videos
09:39Physical, Chemical and Biological Characterization of Six Biochars Produced for the Remediation of Contaminated Sites
Published on: November 28, 2014
11:59Extraction and Detection of Geosmin and 2-Methylisoborneol in Water and Fish using High-Capacity Sorptive Extraction Probes and GC-MS
Published on: July 3, 2025