Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Assessment of a new Goldmann applanation tonometer
1Department of Ophthalmology, University of Bern, Inselspital, Bern 3010, Switzerland.
Background/Aims:
The classic Goldmann applanation tonometer (GAT) has been further developed by Haag-Streit International. The applanation principle has been retained, while the internal force transmission and the pressure gauging have been optimised, the display of results digitised. The authors compared the GAT standard with the new GAT digital.
Methods:
Four fixed tonometer pairs were used. The protocol included: non-contact pachymetry, slit-lamp examination, three consecutive measurements with each tonometer with a 5 min interval in between, check for side effects in 15 min. Three groups (intraocular pressure (IOP) levels) were defined: (1) IOP ≤ 16; (2) IOP>16 and <23; (3) IOP ≥ 23 mm Hg.
Results:
125 Patients (250 eyes) were evaluated. IOP (mm Hg), GAT standard versus GAT digital, for the rights eyes was: Group 1: 12.94 ± 0.55 versus 13.11 ± 0.53, p=0.71. Group 2: 18.26 ± 0.59 versus 18.03 ± 0.52, p=0.53; Group 3: 30.28 ± 0.48 versus 30.42 ± 0.41, p=0.97; all right eyes: 17.48 ± 7.48 versus 17.73 ± 7.4, p=0.99. For the left eyes, there was no significant difference, either. The correlation was very good and was not influenced by the IOP level. The Pearson coefficient for the right eye was 0.985, and for the left eye 0.994. In the Bland-Altman analysis, although there were two single readings that differed by as much as 5 mm Hg, GAT digital measures showed almost no skew, and the mean difference was 0.03 ± 1.23 mm Hg (n=250). A multiple regression analysis showed no influence of order of measurement, eyeside or pachymetry.
Conclusions:
The new GAT digital is as reliable and safe as GAT standard. IOP values correlate well. It offers a digitised display and a wireless transfer of data. The display of values up to the first decimal digit is not necessarily associated with a more precise measurement, but may offer an additional comfort compared with the 2 mm Hg scale of the classic GAT.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation