Voxel-based dose reconstruction for total body irradiation with helical tomotherapy
Ming Chao1, Jose Penagaricano, Yulong Yan
1Department of Radiation Oncology, University of Arkansas for Medical Sciences, Little Rock, Arkansas 72205-7199, USA. mchao@uams.edu
Purpose:
We have developed a megavoltage CT (MVCT)-based dose reconstruction strategy for total body irradiation (TBI) with helical TomoTherapy (HT) using a deformable registration model to account for the patient's interfraction changes. The proposed technique serves as an efficient tool for delivered dose verification and, potentially, plan adaptation.
Methods And Materials:
Four patients with acute myelogenous leukemia treated with TBI using HT were selected for this study. The prescription was 12 Gy, 2 Gy/fraction, twice per day, given at least 6 h apart. The original plan achieved coverage of 80% of the clinical target volume (CTV) by the 12 Gy isodose surface. MVCTs were acquired prior to each treatment. Regions of interest were contoured on each MVCT. The dose for each fraction was calculated based on the MVCT using the HT planned adaptive station. B-spline deformable registration was conducted to establish voxel-to-voxel correspondence between the MVCT and the planning CT. The resultant deformation vector was employed to map the reconstructed dose from each fraction to the same point as the plan dose, and a voxel-to-voxel summed dose from all six fractions was obtained. The reconstructed dose distribution and its dosimetric parameters were compared with those of the original treatment plan.
Results:
While changes in CTV contours occurred in all patients, the reconstructed dose distribution showed that the dose-volume histogram for CTV coverage was close (<1.5%) to that of the original plan. For sensitive structures, the differences between the reconstructed and the planned doses were less than 3.0%.
Conclusion:
Voxel-based dose reconstruction strategy that takes into account interfraction anatomical changes using MVCTs is a powerful tool for treatment verification of the delivered doses. This proposed technique can also be applied to adaptive TBI therapy using HT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...