Successful retrieval of a frayed coronary stent using the peripheral crossover technique
Anand Prasad1, Manjusha Ilapakurti, Amir Ravandi
1Division of Cardiovascular Medicine, University of California San Diego, 200 West Arbor Drive, San Diego, CA 92103, USA. anprasad@ucsd.edu
Abstract:
With the advent of modern thin-strut coronary stents, there is potential for damage to the stent structure during delivery. Fraying of the stent during unsuccessful attempts at lesion passage can be particularly problematic as re-entry into the guiding catheter may not be possible. Familiarity with techniques to retrieve damaged stents is helpful in this situation. In the present case, we describe intentional removal of a frayed stent from its balloon and subsequent retrieval using a peripheral crossover sheath.