Direct detection of uric acid in body fluid samples by using boron doped graphite nano particles as the working
Jianjun He1, Bunshi Fugetsu, Shunitz Tanaka
1Laboratory of Environmental Remediation, Graduate School of Environmental Science, Hokkaido University, Sapporo, Japan.
Abstract:
Boron doped graphite nano particles were used as functioning elements for creation of electrodes for the detection of uric acid in biological samples. The electrode obtained in this manner was capable of oxidizing ascorbic acid at lower potentials. This provided a desirable solution to the interfering problem encountered in the detection of uric acid in biological samples caused by ascorbic acid. The detectable concentrations for uric acid ranged from 5.0 to 130 µM. The applicability of the electrode was experimentally demonstrated by the successful direct detection of uric acid in real urine samples.