Related Experiment Video
Updated: Jun 2, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Positive enteric contrast material for abdominal and pelvic CT with automatic exposure control: what is the effect on
Zhen J Wang1, Katherine S Chen, Robert Gould
1Department of Radiology, University of California San Francisco, 505 Parnassus Avenue, San Francisco, CA 94143-0628, United States. jane.wang@radiology.ucsf.edu
Objective:
To assess the effect of positive enteric contrast administration on automatic exposure control (AEC) CT radiation exposure in (1) a CT phantom, and (2) a retrospective review of patients.
Materials And Methods:
We scanned a CT phantom containing simulated bowel that was sequentially filled with water and positive enteric contrast, and recorded the mean volume CT dose index (CTDIvol). We also identified 17 patients who had undergone 2 technically comparable CT scans of the abdomen and pelvis, one with positive enteric contrast and the other with oral water. Paired Student's t-tests were used to compare the mean CTDIvol between scans performed with and without positive enteric contrast. Both the phantom and patient CT scans were performed using AEC with a fixed noise index.
Results:
The mean CTDIvol for the phantom with simulated bowel containing water and positive enteric contrast were 8.2 ± 0.2 mGy, and 8.7 ± 0.1 mGy (6.1% higher than water, p=0.02), respectively. The mean CTDIvol for patients scanned with oral water and with positive enteric contrast were 11.8 mGy and 13.1 mGy, respectively (p=0.003). This corresponded to a mean CTDIvol which was 11.0% higher (range: 0.0-20.7% higher) in scans with positive enteric contrast than those with oral water in patients.
Conclusions:
When automatic exposure control is utilized for abdominopelvic CT, the radiation exposure, as measured by CTDIvol, is higher for scans performed with positive enteric contrast than those with oral water.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging