Related Experiment Video
Updated: Jun 2, 2026

Randomized Controlled Trial to Study the Acute Effects of Strength Exercise on Insulin Sensitivity in Obese Adults
Published on: December 1, 2023
Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised
Roman Hovorka1, Kavita Kumareswaran, Julie Harris
1Institute of Metabolic Science, University of Cambridge, Cambridge, UK. rh347@cam.ac.uk
Objective:
To compare the safety and efficacy of overnight closed loop delivery of insulin (artificial pancreas) with conventional insulin pump therapy in adults with type 1 diabetes.
Design:
Two sequential, open label, randomised controlled crossover, single centre studies.
Setting:
Clinical research facility.
Participants:
24 adults (10 men, 14 women) with type 1 diabetes, aged 18-65, who had used insulin pump therapy for at least three months: 12 were tested after consuming a medium sized meal and the other 12 after consuming a larger meal accompanied by alcohol.
Intervention:
During overnight closed loop delivery, sensor measurements of glucose were fed into a computer algorithm, which advised on insulin pump infusion rates at 15 minute intervals. During control nights, conventional insulin pump settings were applied. One study compared closed loop delivery of insulin with conventional pump therapy after a medium sized evening meal (60 g of carbohydrates) at 1900, depicting the scenario of "eating in." The other study was carried out after a later large evening meal (100 g of carbohydrates) at 2030, accompanied by white wine (0.75 g/kg ethanol) and depicted the scenario of "eating out."
Main Outcome Measures:
The primary outcome was the time plasma glucose levels were in target (3.91-8.0 mmol/L) during closed loop delivery and a comparable control period. Secondary outcomes included pooled data analysis and time plasma glucose levels were below target (≤ 3.9 mmol/L).
Results:
For the eating in scenario, overnight closed loop delivery of insulin increased the time plasma glucose levels were in target by a median 15% (interquartile range 3-35%), P = 0.002. For the eating out scenario, closed loop delivery increased the time plasma glucose levels were in target by a median 28% (2-39%), P = 0.01. Analysis of pooled data showed that the overall time plasma glucose was in target increased by a median 22% (3-37%) with closed loop delivery (P < 0.001). Closed loop delivery reduced overnight time spent hypoglycaemic (plasma glucose ≤ 3.9 mmol/L) by a median 3% (0-20%), P=0.04, and eliminated plasma glucose concentrations below 3.0 mmol/L after midnight.
Conclusion:
These two small crossover trials suggest that closed loop delivery of insulin may improve overnight control of glucose levels and reduce the risk of nocturnal hypoglycaemia in adults with type 1 diabetes. Trial registration ClinicalTrials.gov NCT00910767 and NCT00944619.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Type I Diabetes I: Introduction
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Type I Diabetes II: Pathophysiology