[Post-traumatic orbital reconstruction comparing preformed versus non preformed titanium mesh plates]
A Momjian1, J Heuberger, P Scolozzi
1Service de chirurgie maxillofaciale et de chirurgie buccale, département de chirurgie, hôpitaux universitaires de Genève, Geneva, Switzerland. Armen.Momjian@hcuge.ch
Revue De Stomatologie Et De Chirurgie Maxillo-Faciale
|April 19, 2011
Summary
Titanium mesh plates for orbital reconstruction showed no significant difference in orbital volume accuracy between non-preshaped (NPMP) and three-dimensionally preformed (PMP) types. Both methods offer reliable orbital volume restoration.
Area of Science:
- Ophthalmology
- Plastic and Reconstructive Surgery
- Biomedical Engineering
Background:
- Orbital reconstruction is crucial for restoring function and aesthetics after trauma or surgery.
- Titanium mesh plates are commonly used, but comparisons between preformed and non-preshaped options are limited.
Purpose of the Study:
- To compare the clinical and radiological accuracy of non-preshaped mesh plates (NPMP) versus three-dimensionally preformed mesh plates (PMP) for orbital reconstruction.
- To evaluate the reliability of each titanium mesh plate type in restoring orbital volume.
Main Methods:
- A comparative study involving 30 patients, with 15 treated using NPMP and 15 using PMP.
- Orbital volumes were calculated from CT scans using OsiriX software, comparing reconstructed and healthy sides.
- Key parameters included sex, age, fracture patterns, and complications like diplopia and enophthalmia.
Main Results:
- No statistically significant difference in orbital volume between reconstructed and contralateral healthy orbits for either NPMP (p=0.991) or PMP (p=0.200) groups.
- Mean orbital volumes of reconstructed orbits were comparable between NPMP (21.76 cm³) and PMP (20.28 cm³) groups (p=0.441).
Conclusions:
- Both non-preshaped and three-dimensionally preformed titanium mesh plates yield comparable results in orbital volume reconstruction.
- The choice between NPMP and PMP may not significantly impact volumetric accuracy in orbital reconstructions based on this study.