Turbulence characteristics downstream of a new trileaflet mechanical heart valve
Chi-Pei Li1, Sheng-Fu Chen, Chi-Wen Lo
1Department of Water Resources and Environmental Engineering, Tamkang University, Taipei, Taiwan.
Abstract:
Design limitations of current mechanical heart valves cause blood flow to separate at the leaflet edges and annular valve base, forming downstream vortex mixing and high turbulent shear stresses. The closing behavior of a bileaflet valve is associated with reverse flow and may lead to cavitation phenomenon. The new trileaflet (TRI) design opens similar to a physiologic valve with central flow and closes primarily due to the vortices in the aortic sinus. In this study, we measured the St. Jude Medical 27 mm and the TRI 27 mm valves in the aortic position of a pulsatile circulatory mock loop under physiologic conditions with digital particle image velocimetry (DPIV). Our results showed the major principal Reynolds shear stresses were <100 N/m2 for both valves, and turbulent viscous shear stresses were smaller than 15 N/m2. The TRI valve closed more slowly than the St. Jude Medical valve. As the magnitudes of the shear stresses were similar, the closing velocity of the valves should be considered as an important factor and might reduce the risks of thrombosis and thromboembolism.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Regurgitation I: Introduction
Aortic Regurgitation I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Mitral Valve Prolapse I: Introduction
Mitral Stenosis I: Introduction