Comparison between Tonopachy and other tonometric and pachymetric devices
Yong Gi Lee1, Ji Hyun Kim, Na Rae Kim
1Department of Ophthalmology, Institute of Vision Research, Yonsei University College of Medicine, Seoul, Korea.
Purpose:
This study aimed to compare central corneal thickness (CCT) measurement by Tonopachy to that by Pentacam and ultrasound pachymetry, and intraocular pressure (IOP) measurement to that by Goldmann applanation tonometry (GAT). The reproducibility of CCT and IOP measurements by Tonopachy was also evaluated.
Methods:
In 104 eyes of 104 patients, CCT was measured by Tonopachy, Pentacam, and ultrasound pachymetry, and IOP was measured by Tonopachy and GAT. Each CCT and IOP measurement was compared using Pearson correlation, repeated measures analysis of variance, and Bland-Altman plots. In 30 subjects, CCT and IOP measurements by Tonopachy were repeated to evaluate intrasession and intraobserver variability.
Results:
Both CCT and IOP measurements were highly correlated among all instruments used in this study. CCT measurements by three pachymeters were statistically different (p<0.0001). CCT measured by ultrasound pachymetry was lowest (541.7 ± 30.6 μm) whereas those by Tonopachy and Pentacam showed no difference (557.3 ± 34.3 and 558.0 ± 33.7 μm, respectively). Tonopachy overestimated CCT by 13.9 μm when compared with ultrasound pachymetry. There was a statistically significant difference between IOP measurements by two tonometers (p<0.0001); IOP measurement was higher by Tonopachy than by GAT (13.9 ± 4.2 and 12.5 ± 3.2 mm Hg, respectively). Tonopachy overestimated IOP measurements by 1.2 mm Hg compared with GAT. Intersession agreements for IOP and CCT measurements by Tonopachy were excellent (intraclass correlation, 0.902 and 0.962, respectively) with 95% limits of agreement ranging from -1.4 to 2.2 mm Hg and from -12.9 to 12.1 μm, respectively.
Conclusions:
Although CCT and IOP measurements obtained by Tonopachy were reproducible and showed close agreement with ultrasound pachymetry and Pentacam, and GAT, careful attention should be paid when comparing Tonopachy CCT measurement to ultrasound pachymetry or its IOP measurement to GAT as the values may not be interchangeable. Tonopachy is a reliable instrument for evaluating CCT and IOP.
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Potentiometric Titration: Overview
Methods of Obtaining Topography
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...