Development of Multiplex Real-Time RT-qPCR Assays for the Detection of SARS-CoV-2, Influenza A/B, and MERS-CoV
Published on: November 10, 2023
Comparison of two multiplex methods for detection of respiratory viruses: FilmArray RP and xTAG RVP
Kenneth H Rand1, Howard Rampersaud, Herbert J Houck
1Department of Pathology, Immunology and Laboratory Medicine, University of Florida, Gainesville, FL 32610, USA. rand@pathology.ufl.edu
Abstract:
We compared the FilmArray RP (Idaho Technology, Inc., Salt Lake City, UT) and the xTAG RVP (Luminex Corporation, Toronto, Canada) multiplex respiratory virus PCR methods for the detection of respiratory viruses in a set of 200 patient specimens frozen at -70 °C after standard viral culture and antigen detection methods were done. Both systems detected between 40 to 50% more viruses than traditional methods, primarily rhinoviruses and human metapneumovirus. The FilmArray RP detected significantly more total viruses either alone or as part of mixed infections than the xTAG RVP, as well as an additional 21.6% more respiratory syncytial viruses. The xTAG RVP requires 5 to 6 h with 2.5 to 3 h of hands-on time, while the FilmArray RP takes about an hour with 3 to 5 min of hands-on time, making it much easier to perform.