Is metal-on-metal squeaking related to acetabular angle of inclination?
Thomas Bernasek1, David Fisher, David Dalury
1Florida Orthopaedic Institute, 13020 Telecom Parkway, N Tampa, FL 33637, USA. tbernasek@floridaortho.com
Background:
Postoperative audible squeaking has been well documented in ceramic-on-ceramic hip prostheses, and several metal-on-metal (MOM) THA designs, specifically those used for large-head resurfacing and MOM polyethylene sandwich designs, and are attributed to different implant- and patient-specific factors. Current literature does not identify the incidence of squeaking in modular MOM THA or possible etiologic factors.
Questions/Purposes:
Our purposes were to (1) identify the incidence of squeaking in modular MOM prostheses in THA; (2) determine whether males or females were more likely to have squeaking; and (3) determine whether the incidence of squeaking relates to acetabular inclination angle.
Methods:
We retrospectively reviewed the patient records and radiographs of 539 patients (542 hips) from three independent centers who underwent a MOM THA between February 2001 and December 2005. Demographic and implant factors were evaluated, including measurement of cup inclination angles. The minimum followup was 36 months (mean, 76 months; range, 36-119 months).
Results:
We identified squeaking in eight of the 542 hips (1.5%); five were in women and two were in men (one patient had bilateral squeaking). The time to onset of patient-reported audible squeaking averaged 23 months (range, 6-84 months). Squeaking was more likely to occur in women (six of eight hips). No hips with 45º or less acetabular inclination squeaked (291 hips); eight of 251 hips (3.2%) with inclination angles greater than 45º squeaked. Patients who reported squeaking had higher inclination angles than those who did not report squeaking.
Conclusions:
Our observations suggest an increased frequency of squeaking in female patients and in patients with greater inclination angles with this MOM implant design.