Ultrafast Time-resolved Near-IR Stimulated Raman Measurements of Functional π-conjugate Systems
Published on: February 10, 2020
Raman spectroscopy using 1550 nm (retina-safe) laser excitation
Carl Brouillette1, Hermes Huang, Wayne Smith
1Real-Time Analyzers, Inc., Middletown, Connecticut 06457, USA.
Abstract:
During the past decade, the use of portable Raman analyzers for field measurements has grown dramatically. However, most analyzers use 785 nm excitation lasers that can cause permanent eye damage. To overcome this safety concern, we have built a portable Fourier transform (FT) Raman analyzer using a 1550 nm retina-safe excitation laser and have compared its performance to our 1064 nm FT-Raman analyzer, which uses the same optical design. Raman theory predicts approximately five times lower peak intensities at 1550 nm. Although we found that intensities were as much as 20 times less intense, the analyzer is still capable of measuring spectra of sufficient quality to identify and differentiate chemicals.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...