Comparing the digoxin test with the Architect® i1000sr system with respect to the AxSYM® system
E Albert Vicent1, R Ferriols Lisart, M A Roch Ventura
1Servicio de Farmacia, Hospital General de Castellón, Castellón, Spain.
Summary
The Architect® system shows precise digoxin measurements but is clinically inaccurate compared to the AsSYM® assay. This impacts patient monitoring due to significant concentration differences.
Area of Science:
- Clinical Chemistry
- Immunoassay Technology
- Therapeutic Drug Monitoring
Background:
- Digoxin is a critical medication requiring precise therapeutic drug monitoring.
- Existing assays like the Digoxin II on AsSYM® provide a benchmark for digoxin level determination.
- Newer automated analyzers require validation against established methods.
Purpose of the Study:
- To evaluate the performance of the Architect® i1000sr analyzer for serum digoxin determination.
- To compare the Architect® digoxin assay against the established AsSYM® Digoxin II assay.
- To assess the clinical utility of the Architect® system for digoxin monitoring.
Main Methods:
- Prospective analysis of 100 patient serum samples undergoing digoxin therapy.
- Simultaneous processing of samples on both AxSYM® and Architect® platforms.
- Statistical assessment using linear regression, coefficient of determination, mean absolute error, mean squared prediction error, and Bland-Altman analysis.
Main Results:
- A high correlation coefficient (0.93) was observed between the two systems.
- Significant discrepancies noted: nearly 40% difference between 0.8-2 ng/ml and 20% in other ranges.
- The Architect® system demonstrated high precision in its measurements.
Conclusions:
- The Architect® system exhibits high precision for digoxin measurement.
- However, the system's accuracy is clinically unacceptable for patient monitoring when compared to the AxSYM® assay.
- Discrepancies may affect therapeutic drug management decisions.
Related Concept Videos
Instrumentation Amplifier
An electrocardiography (ECG) machine is an essential piece of medical equipment used to monitor the electrical activity of the heart. It operates by detecting small electrical changes on the skin that result from the depolarization of the heart muscle during each heartbeat. However, these signals are in the microvolt range and can be easily overwhelmed by noise or interference.
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Automated Microbial Diagnostics
Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...
Equipments Used To Measure Blood Pressure
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias, also known as arrhythmias, are disturbances in the heart's rhythm that range from benign to life-threatening. A thorough evaluation is crucial for appropriate management and involves a comprehensive medical history, physical examination, and various diagnostic tests.Medical HistorySymptoms: Collect detailed information on palpitations, dizziness, syncope, chest pain, and fatigue. Note their onset, frequency, and triggers.Previous Cardiac Issues: Document any history of heart...