Quantitative computed tomography assessment of transfusional iron overload
John C Wood1, Ashley Mo, Aakansha Gera
1Division of Paediatric Cardiology, Department of Paediatrics, Children's Hospital Los Angeles, CA 90027, USA. jwood@chla.usc.edu
Quantitative computed tomography (QCT) shows promise for measuring liver iron concentration (LIC) in patients with transfusional siderosis, especially when MRI is unavailable. QCT is viable for moderate to severe iron levels but not for well-controlled cases.
Area of Science:
- Radiology
- Medical Imaging
- Biomedical Engineering
Background:
- Quantitative computed tomography (QCT) has been explored for liver iron quantification for over three decades.
- Clinical validation of QCT for liver iron concentration (LIC) remains limited.
- Transfusional siderosis necessitates accurate monitoring of iron overload.
Purpose of the Study:
- To compare liver attenuation measurements from QCT with magnetic resonance imaging (MRI)-derived LIC estimates.
- To assess the clinical utility of QCT for quantifying liver iron in patients with transfusional siderosis.
Main Methods:
- A cohort of 37 patients with transfusional siderosis underwent both QCT and MRI for LIC monitoring.
- QCT and MRI measurements were performed within a 6-year period, with a mean time difference of 14 days.
- Liver attenuation values from QCT were correlated with MRI-derived LIC.
Main Results:
- A linear relationship was observed between liver attenuation and LIC for values outside the normal range (r(2) = 0.94).
- Intersubject variability in intrinsic liver attenuation limited QCT's ability to quantify LIC below 8 mg/g dry weight.
- QCT and MRI demonstrated equivalent accuracy for LIC around 22 mg/g dry weight, with QCT outperforming MRI at higher LIC levels.
Conclusions:
- QCT is not suitable for monitoring patients with well-controlled iron levels due to measurement uncertainty.
- QCT presents a cost-effective, available, and high-throughput alternative for liver iron quantification in moderate to severe cases, particularly where MRI resources are scarce.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Imaging Studies for Cardiovascular System V: CT