Validation of a microbial detection system for use with ACD-A platelets with PAS III platelet additive solution
Larry J Dumont1, Shauna N Hay, Louise Herschel
1Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire 03756, USA. larry.j.dumont@hitchcock.org
Transfusion
|April 27, 2011
Summary
The two-bottle BacT/ALERT microbial detection system (BTA) effectively detects microorganisms in leukoreduced apheresis platelets (LR-AP) with InterSol. This method shows superior performance compared to traditional plate cultures for platelet safety.
Area of Science:
- Microbiology
- Transfusion Medicine
- Clinical Diagnostics
Background:
- The BacT/ALERT microbial detection system (BTA) is utilized for testing leukoreduced apheresis platelets (LR-AP) in plasma.
- Platelet additive solutions (PASs), such as InterSol, are employed to minimize plasma transfusion in LR-AP.
Purpose of the Study:
- To evaluate the efficacy of the two-bottle BTA testing scheme for microorganism recovery from LR-AP in InterSol-plasma.
- To compare the BTA system's performance against a reference plate culture method.
Main Methods:
- Hyperconcentrated, double LR-AP were collected and mixed with InterSol (65% InterSol:35% plasma).
- Samples were inoculated with one of 10 index organisms at two levels and tested using aerobic (BPA) and anaerobic (BPN) BTA bottles (4 mL each).
- A reference method used blood agar plates (BAPs) for aerobic and anaerobic cultures (0.5 mL each).
Main Results:
- The BTA system demonstrated zero false-positives from 103 bottle pairs.
- All 400 two-bottle BTA tests were positive within 24 hours, with exceptions for Propionibacterium acnes and Streptococcus viridans.
- At bacterial concentrations of 40 colony-forming units (CFUs)/mL or less, BTA achieved 100% positivity, compared to 94% for BAP.
Conclusions:
- The two-bottle BTA system exhibited superior seeded organism recovery compared to the two-plate BAP system for InterSol platelets.
- The performance of the BTA system with InterSol platelets is comparable to previous findings with platelets in plasma.
- InterSol additive solution does not appear to negatively impact the performance of the two-bottle BTA detection system.