Prospective evaluation of the self-pressurized air-Q intubating laryngeal airway in children
Narasimhan Jagannathan1, Lisa E Sohn, Ravinder Mankoo
1Department of Pediatric Anesthesiology, Children's Memorial Hospital, Northwestern University Feinberg School of Medicine, Chicago, IL 60614, USA. simjag2000@yahoo.com
Objectives:
To assess the clinical efficacy of the self-pressurized air-Q ILA™ (ILA-SP).
Aim:
The purpose of this prospective audit was to evaluate the feasibility of the ILA-SP in clinical practice and generate data for future comparison trials.
Background:
The ILA-SP is a new first-generation supraglottic airway for children with a self-adjusting cuff and lack of a pilot balloon.
Methods:
Over a 4-month period, 352 children with an ASA physical status of I-III, newborn to 18 years of age, undergoing various procedures were studied. Data points assessed included insertion success rates, airway leak pressures, quality of ventilation, and perioperative complications associated with the use of this device.
Results:
In 349 of the 352 patients in this study, the ILA-SP was used successfully as a primary supraglottic airway device in a variety of patients. Three patients required conversion to a standard laryngeal mask airway or a tracheal tube. The mean initial airway leak pressure for all patients was 17.8 ± 5.4 cm H(2)O, and 20.4 ± 5.5 cm H(2)O when re-checked at 10 min, which was statistically significant (P < 0.001). Complications were limited to 14 patients and related to reflex activation of the airway (coughing, laryngospasm, and bronchospasm) (n = 10), sore throat (n = 3), and blood staining on removal of the device (n = 1). There were no episodes of regurgitation, aspiration, or hoarseness.
Conclusions:
Acceptable clinical performance was demonstrated with the ILA-SP for a variety of procedures in infants and children with spontaneous and positive pressure ventilation. Future studies comparing this device to other supraglottic airways may provide useful information regarding the safety of the ILA-SP in pediatric clinical practice.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...