The effect of ventilator performance on airway pressure release ventilation: a model lung study
Takeshi Yoshida1, Akinori Uchiyama, Takashi Mashimo
1Department of Anesthesiology and Intensive Care Medicine, Osaka University Graduate School of Medicine, 2-15 Yamadaoka, Suita, Osaka, Japan 565-0871. doctor@mvj.biglobe.ne.jp
Background:
Using a model lung connected to six different ventilators, with each ventilator in the airway pressure release ventilation mode, we measured differences in intrinsic positive end-expiratory pressure (PEEPi) during the expiratory phase and calculated the inspiratory and expiratory pressure time product (PTP) as an index of work of breathing during the inspiratory phase.
Methods:
We compared 6 ventilators: Puritan-Bennett 840, Evita XL, Servo i, Avea, Hamilton G5, and Engström. With a constant inspiratory pressure level of 25 cm H(2)O and expiratory pressure level of 0 cm H(2)O, PEEPi was measured as the expiratory time was decremented from 1.0 second to 0.2 second in steps of 0.1 second. The inspiratory and expiratory PTPs were measured during the ventilator's inspiratory phase by simulating spontaneous breathing with a tidal volume of 300 mL, with a respiratory rate of 30 breaths/min and with expiratory flow rates of 0.5 L/s, 1.0 L/s, and 1.5 L/s.
Results:
In all ventilators, the progressive diminution of the expiratory time caused a significant increase in PEEPi (P< 0.001). With a 0.2-second expiratory time, PEEPi ranged from 9.4± 0.07 cm H(2)O for the Servo i to 15.7± 0.04 cm H(2)O for the Avea. The Servo i had a significantly lower inspiratory PTP than did the other ventilators (P< 0.001). When the expiratory flow rate was 0.5 L/s and 1.0 L/s, the expiratory PTP was lower with the Servo i and Evita XL than with the other ventilators (P< 0.001).
Conclusions:
PEEPi varied significantly among ventilators. Inspiratory and expiratory work of breathing varied between ventilators when spontaneous breathing occurred during the ventilator's inspiratory phase.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)