Related Experiment Video
Updated: Jun 2, 2026

Transcutaneous Assessment of Renal Function in Conscious Rodents
Published on: March 26, 2016
Assessing renal function with a rapid, handy, point-of-care whole blood creatinine meter before using contrast
Satoru Morita1, Kazufumi Suzuki, Ai Masukawa
1Department of Radiology, Tokyo Women's Medical University, Medical Center East, 2-1-10 Nishiogu, Arakawa-ku, Tokyo 116-8567, Japan. i@imodey.com
Purpose:
The aim of this prospective study was to assess the reliability of a rapid, handy, point-of-care whole blood creatinine meter (PCM) in patients who were scheduled to undergo contrast-enhanced computed tomography (CT) or magnetic resonance imaging (MRI).
Materials And Methods:
Among patients scheduled to undergo contrast-enhanced CT or MRI examinations, 113 patients who did not have creatinine data from the prescribed intervals before the examination day (in principle, 90 days for scheduled outpatients and 7 days for inpatients and urgent patients) were included. The estimated glomerular filtration rate (eGFR) was calculated using the creatinine values measured with the PCM and those from central laboratory measurements (LAB). The two eGFR values were compared statistically with the paired t-test, Pearson's correlation coefficient, and the Bland-Altman analysis.
Results:
The mean eGFR measured with the PCM was slightly higher than the LAB value (81.2 ± 24.6 vs. 70.2 ± 19.7 ml/min/1.73 m(2), P < 0.0001). They were well correlated (r = 0.74, y = 0.92x + 16.9, P < 0.0001). Bland-Altman plots (mean difference was 11.0 ml/min/1.73 m(2); limits of agreement were -22.4 to +44.4 ml/min/1.73 m(2)) showed a moderate agreement with some degree of dispersion.
Conclusion:
The PCM can rapidly assess renal function using a small amount of blood almost equally to that of determined in the laboratory, which may help reduce the risk of contrast material-induced complications.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...