Related Experiment Video
Updated: Jun 2, 2026

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
[Stricture of oxygen outlet of the central piping identified by a decrease in the oxygen supply pressure into the
Yoko Desaki1, Toshihiro Yorozuya, Kazuo Nakanishi
1Department of Anesthesiology, Uwajima City Hospital, Uwajima 798-8510.
Degradation of an O-ring in medical oxygen lines caused intermittent pressure drops. Monitoring pressure fluctuations during high-flow oxygen use is crucial for detecting this serious anesthesia machine malfunction.
Area of Science:
- Biomedical Engineering
- Medical Device Safety
Context:
- Anesthesia delivery systems rely on consistent oxygen supply.
- Intermittent pressure drops in central piping oxygen outlets can go unnoticed during routine checks.
- Previous similar incidents indicate a recurring issue with O-ring degradation.
Purpose:
- To report an incident of oxygen supply pressure malfunction in an anesthesia machine.
- To highlight the importance of monitoring pressure deviations during high-flow oxygen use.
- To identify O-ring degradation as a cause of intermittent oxygen supply failure.
Summary:
- An O-ring degradation in the central piping oxygen outlet led to intermittent decreases in oxygen supply pressure to the anesthesia machine.
- The malfunction was initially missed due to periodic normal pressure readings.
- Cyclical pressure drops correlated with increased oxygen consumption, signaling a critical failure.
Impact:
- Emphasizes the need for vigilant monitoring of oxygen supply pressure, especially under high-flow conditions.
- Suggests that O-ring degradation is a potential cause of critical anesthesia equipment failure.
- Highlights the limitations of periodic checks in detecting dynamic pressure fluctuations.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.