Patient radiation dose audits for fluoroscopically guided interventional procedures
Stephen Baiter1, Marvin Rosenstein, Donald L Miller
1Department of Radiology, Columbia University, New York, New York 10032, USA. sb2455@columbia.edu
Purpose:
Quality management for any use of medical x-ray imaging should include monitoring of radiation dose. Fluoroscopically guided interventional (FGI) procedures are inherently clinically variable and have the potential for inducing deterministic injuries in patients. The use of a conventional diagnostic reference level is not appropriate for FGI procedures. A similar but more detailed quality process for management of radiation dose in FGI procedures is described.
Methods:
A method that takes into account both the inherent variability of FGI procedures and the risk of deterministic injuries from these procedures is suggested. The substantial radiation dose level (SRDL) is an absolute action level (with regard to patient follow-up) below which skin injury is highly unlikely and above which skin injury is possible. The quality process for FGI procedures collects data from all instances of a given procedure from a number of facilities into an advisory data set (ADS). An individual facility collects a facility data set (FDS) comprised of all instances of the same procedure at that facility. The individual FDS is then compared to the multifacility ADS with regard to the overall shape of the dose distributions and the percent of instances in both the ADS and the FDS that exceed the SRDL.
Results:
Samples of an ADS and FDS for percutaneous coronary intervention, using the dose metric of reference air kerma (K(a,r)) (i.e., the cumulative air kerma at the reference point), are used to illustrate the proposed quality process for FGI procedures. Investigation is warranted whenever the FDS is noticeably different from the ADS for the specific FGI procedure and particularly in two circumstances: (1) When the facility's local median K(a,r) exceeds the 75th percentile of the ADS and (2) when the percent of instances where K(a,r) exceeds the facility-selected SRDL is greater for the FDS than for the ADS.
Conclusions:
Analysis of the two data sets (ADS and FDS) and of the percent of instances that exceed the SRDL provides a means for the facility to better manage radiation dose (and therefore both deterministic and stochastic radiation risk) to the patient during FGI procedures.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...