Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Extremely sensitive detection of NO₂ employing off-axis integrated cavity output spectroscopy coupled with
Gottipaty N Rao1, Andreas Karpf
1Department of Physics, Adelphi University, Garden City, New York 11530, USA. Rao@adelphi.edu
Abstract:
We report on the development of a new sensor for NO₂ with ultrahigh sensitivity of detection. This has been accomplished by combining off-axis integrated cavity output spectroscopy (OA-ICOS) (which can provide large path lengths of the order of several kilometers in a small volume cell) with multiple-line integrated absorption spectroscopy (MLIAS) (where we integrate the absorption spectra over a large number of rotational-vibrational transitions of the molecular species to further improve the sensitivity). Employing an external cavity quantum cascade laser operating in the 1601-1670 cm⁻¹ range and a high-finesse optical cavity, the absorption spectra of NO₂ over 100 transitions in the R band have been recorded. From the observed linear relationship between the integrated absorption versus concentration of NO₂ and the standard deviation of the integrated absorption signal, we report an effective sensitivity of detection of approximately 28 ppt (parts in 10¹²) for NO₂ To the best of our knowledge, this is among the most sensitive levels of detection of NO₂ to date.
Related Concept Videos
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Instrumentation
Molecular Spectroscopy: Absorption and Emission
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...