Atrial oversensing by an LV lead during typical flutter: the interest of electronic repositioning
Alexandre Duparc1, Pierre Mondoly, Nicolas Detis
1Unité rythmologie et stimulation cardiaque, Pôle Cardiovasculaire et Métabolique, University Hospital of Rangueil, Toulouse, France. duparc.a@chu-toulouse.fr
Abstract:
Occurrence of supraventricular tachycardia is a common cause of clinical impairment for patients implanted with CRT devices. We report the case of atrial activity oversensing by the left ventricular (LV) lead during typical flutter, which led to LV pacing inhibition. Temporary reprogramming of the LV detection from standard bipolar to extended bipolar and cavotricuspid isthmus ablation solved this problem.