Related Experiment Video
Updated: Jun 2, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
Determination of pethidine hydrochloride using potentiometric coated graphite and carbon paste electrodes
Hazem M Abu-Shawish1, Ayoub A Dalou, Nasser Abu Ghalwa
1Faculty of Science, Research Center, Al-Aqsa University, Gaza, Palestine. hazemona1@yahoo.co.uk
Abstract:
A new approach for lowering the detection limit of a pethidine ion-selective electrode is presented. A coated graphite (CGE) and carbon paste (CPE) electrodes for pethidine ions based on pethidine-phosphotungstate (PD-PT) as ion-pair complex are described. The sensors exhibit a Nernstian slope of 58.1 and 54.2 mVdecade(-1) for pethidine ion over a wide concentration range from 2.6 × 10(-7) to 1.0 × 10(-2) M and 2.1 × 10(-6) to 1.0 × 10(-2) M with a detection limit of 1.8 × 10(-7) M and 7.3 × 10(-7) M for pethidine coated graphite (PD-CGE) and pethidine carbon paste electrode (PD-CPE), respectively. These sensors exhibited a fast response time (about 5-8 s) and good stability. The standard electrode potentials, E(o) , were determined at different temperatures and used to calculate the isothermal temperature coefficient (dE(o) /dT) of the PD-CGE and PD-CPE, which was 0.0062 and 0.0071 V/ °C, respectively. Selectivity coefficients, determined by matched potential method (MPM) and separate solution method (SSM), showed high selectivity for pethidine hydrochloride (PDCl) over a large number of inorganic cations, organic cations, sugars, urine components, and some common drug excipients. The sensors were applied for determination of PDCl in ampoule and in spiked urine samples using potentiometric determination, standard addition and the calibration curve methods. The results obtained were satisfactory with excellent percentage recovery comparable and sometimes better than those obtained by other routine methods for the assay.
More Related Videos
10:48Electrochemical Preparation of Poly(3,4-Ethylenedioxythiophene) Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
11:18Manufacturing of a Nafion-coated, Reduced Graphene Oxide/Polyaniline Chemiresistive Sensor to Monitor pH in Real-time During Microbial Fermentation
Published on: January 7, 2019
Related Concept Videos
Potentiometry: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Potentiometric Titration: Overview
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Voltammetric Techniques: Linear-Scan (E vs Time)