Related Experiment Video
Updated: Jun 2, 2026

Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Activated charcoal effectively removes inhaled anesthetics from modern anesthesia machines
Nathaniel Birgenheier1, Robert Stoker, Dwayne Westenskow
1University of Utah, Salt Lake City, UT 84132, USA. nathaniel.birgenheier@hsc.utah.edu
Introduction:
If a malignant hyperthermia-susceptible patient is to receive an anesthetic, an anesthesia machine that has been used previously to deliver volatile anesthetics should be flushed with a high fresh gas flow. Conflicting results from previous studies recommend flush times that vary from 10 to 104 minutes. In a previously proposed alternative decontamination technique, other investigators placed an activated charcoal filter in the inspired limb of the breathing circuit.
Methods:
We placed activated charcoal filters on both the inspired and expired limbs of several contaminated anesthesia machines and measured the time needed to flush the machine so that the delivered concentrations of isoflurane, sevoflurane, and desflurane would be <5 parts per million (ppm). We next simulated the case for which malignant hyperthermia is diagnosed 90 minutes after induction of anesthesia and measured how well activated charcoal filters limit further exposure.
Results:
Activated charcoal filters decrease the concentration of volatile anesthetic delivered by a contaminated machine to an acceptable level in <2 minutes. The concentrations remained well below 5 ppm for at least 60 minutes. When malignant hyperthermia is diagnosed after induction of anesthesia, we found that with charcoal filters in place, the current anesthesia machine may be used for at least 67 minutes before the inspired concentration exceeds 5 ppm.
Conclusions:
Activated charcoal filters provide an alternative approach to the 10 to 104 minutes of flushing that are normally required to prepare a machine that has been used previously to deliver a volatile anesthetic.
Related Concept Videos
Inhalational Anesthetics: Overview
Prevention of Further Absorption of Poison
Anticholinesterase Agents: Poisoning and Treatment
Irreversible agents form a strong bond with the cholinesterase enzyme, making it inactive. The breakdown of the phosphorylated enzyme is slower than the...
Skeletal Muscle Relaxants: Therapeutic Uses
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Stages of General Anesthesia