Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station
Marlin Wayne Causey1, Seth Miller, Andrew Foster
1Department of Surgery, Madigan Army Medical Center, Tacoma, WA 98431, USA. mwcausey@msn.com
Background:
Hemoglobin levels must be obtained through blood draws, which are invasive, time-consuming, and provide only 1 data point at a time rather than continuous measurements. The Masimo Radical-7 SpHb Station (Masimo Corporation, Irvine, CA) has been shown by its manufacturers to provide accurate noninvasive hemoglobin measurements in physiologically normal patients. The objective of this study was to validate noninvasive hemoglobin measurements using the Masimo Radical-7 device.
Methods:
Data were prospectively collected in 2 cohorts of patients: major operations requiring hemodynamic monitoring (operating room [OR]) and critically ill patients (intensive care unit [ICU]). Noninvasive hemoglobin measurements (SpHb) were recorded and were then compared with laboratory hemoglobin measurements.
Results:
Data were collected on 60 patients (OR = 25 and ICU = 45). The overall correlation of the Masimo SpHb and the laboratory Hb was .77 (P < .001) in the OR group with a mean difference of .29 g/dL (95% confidence interval [CI], .08-.49). The overall correlation in the ICU group was .67 (P < .001) with a mean difference of .05 g/dL (95% CI, -.22 to -.31).
Conclusions:
Noninvasive hemoglobin monitoring is a new technology that correlated with laboratory values and supports the continued study of noninvasive hemoglobin monitoring.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: