Related Experiment Video
Updated: Jun 2, 2026

Experimental Protocol for Detecting Cyanobacteria in Liquid and Solid Samples with an Antibody Microarray Chip
Published on: February 7, 2017
Near real-time, on-site, quantitative analysis of PAHs in the aqueous environment using an antibody-based biosensor
Candace R Spier1, George G Vadas, Stephen L Kaattari
1Virginia Institute of Marine Science, College of William & Mary, Gloucester Point, Virginia, USA.
Abstract:
Rapid, on-site, quantitative assessments of dissolved polycyclic aromatic hydrocarbons (PAHs) were demonstrated for two field applications. The platform, a KinExA Inline Sensor (Sapidyne Instruments), employed the monoclonal anti-PAH antibody, 7B2.3, which has specificity for 3- to 5-ring PAHs. A spatial study was conducted near a dredging site where contaminated sediments were being removed, and a temporal study was performed during a rainfall event. Most importantly, the generation of near real-time data guided management decisions in the field and determined proper sampling protocols for conventional analyses. The method was able to determine PAH concentrations as low as 0.3 µg/L, within 10 min of sample acquisition, and to assess 80+ samples (not including standards and blanks) in less than 3 d. These results were compared with a laboratory-based gas chromatography-mass spectrometry method in which a wide array of PAHs, including alkylated homologs, were examined. This system shows great promise as a field instrument for the rapid monitoring of PAH pollution.