Evaluation of methods for preserving PTEN antigenicity in stored paraffin sections

Arnold B Gelb1, Vicki A Freeman, Stephanie H Astrow

  • 1Biomarker Anatomic Pathology Laboratory, Quest Diagnostics Clinical Trials, Valencia, CA, USA. arnold.b.gelb@questdiagnostics.com

Insights

Preserving phosphatase and tensin homolog (PTEN) antigenicity in breast cancer specimens is crucial for accurate assessment. Refrigerated storage with paraffin coating offers optimal PTEN signal preservation for up to four months.

Area of Science:

  • Oncology
  • Biochemistry
  • Pathology

Background:

  • Reduced phosphatase and tensin homolog (PTEN) expression is linked to phosphatidylinositol 3 kinase/Akt pathway inhibitor efficacy in breast cancer.
  • Accurate assessment of PTEN levels is vital, but antigen preservation can be compromised during sample storage.
  • Minimizing signal loss due to poor antigen preservation, rather than actual protein decrease, is a key challenge.

Purpose of the Study:

  • To evaluate methods for preserving PTEN antigenicity in breast cancer tissue sections.
  • To compare the impact of paraffin coating and refrigerated storage on PTEN immunoreactivity over time.

Main Methods:

  • Semiquantitative assessment of PTEN immunoreactivity (histoscore) in 26 breast cancer specimens.
  • Comparison of freshly cut sections with sections stored under various conditions: paraffin-coated vs. uncoated, room temperature (RT) vs. refrigerated (2-8°C) for up to 6 months.

Main Results:

  • Freshly cut sections showed the strongest PTEN histoscore.
  • Refrigerated paraffin-coated sections maintained the highest PTEN signal among stored samples, followed by refrigerated uncoated sections.
  • Room temperature storage, especially without paraffin coating, resulted in significant PTEN signal loss (up to 75% at 4 months).

Conclusions:

  • While freshly cut sections are optimal for PTEN assessment, refrigerated storage with paraffin coating preserves antigenicity for up to 4 months.
  • Refrigeration alone (with or without paraffin) provides adequate PTEN antigen preservation for at least 2 months.
  • Proper sample handling and storage are critical to avoid misinterpreting antigen loss as reduced PTEN expression.