Accuracy of automated community pharmacy-based blood pressure devices
Jasmine Gonzalvo1, Alan Zillich
1College of Pharmacy, Purdue University, Lafayette, IN, USA. jgonzalv@purdue.edu
Objectives:
To estimate accuracy and reliability of pharmacy-based fixed-location automated blood pressure devices (ABPDs) and to test the hypothesis that an ABPD is less accurate with more variable results than a home blood pressure device (HBPD).
Methods:
Randomized study comparing 99 ABPDs with an Omron Digital HBPD in Indiana pharmacies. Each site was visited by one of five study investigators. A questionnaire was used to collect information about ABPDs. To test the ABPD against the HBPD, investigators measured their own blood pressure with each device three times in random order.
Results:
No significant differences were observed between HBPD and ABPD diastolic readings, whereas a statistically significant difference between HBPD and ABPD systolic readings was found. ABPD measurements are as reliable as HBPD measurements when comparing single measurements from each, but reliability differs with more than one reading.
Conclusion:
Compared with a valid HBPD, the ABPD produces inaccurate systolic blood pressure values but similar reliability. Regular blood pressure measurement by health professionals remains optimal for managing hypertensive individuals.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.