Effect of gel-instillation sonography on Doppler ultrasound findings in endometrial polyps
T Van Den Bosch1, D Van Schoubroeck, J Luts
1Department of Obstetrics & Gynecology, University Hospitals KU Leuven, Leuven, Belgium. thierry.van.den.bosch@skynet.be
Objective:
Saline infusion sonohysterography has been reported to suppress the color signal within the endometrium at color or power Doppler evaluation. The aim of this study was to evaluate if gel-instillation sonography (GIS) affects the power Doppler signal in patients with endometrial polyps.
Methods:
Ultrasound volumes of the uterus, obtained by three-dimensional ultrasound imaging of 25 women with histologically confirmed endometrial polyps, were assessed offline by six gynecologists with a special interest in gynecological ultrasound. Each woman contributed four volumes: one gray-scale volume and one power Doppler volume before GIS, and one gray-scale volume and one power Doppler volume at GIS. Power Doppler features before and after gel infusion were compared.
Results:
At unenhanced ultrasound a pedicle artery was seen in 27-46% of cases, whereas, after gel infusion the examiners reported a pedicle artery in 30-46% of cases (Exact McNemar's test P-values ranged from 0.50 to 1.00). The level of agreement between unenhanced ultrasound and GIS ranged from 59 to 91% (Cohen's kappa values ranged from 0.17 to 0.79). There was no tendency for a pedicle artery to be identified less often at GIS than before gel instillation.
Conclusion:
Gel infusion does not affect the power Doppler signal in patients with endometrial polyps.