Estimating central SBP from the peripheral pulse: influence of waveform analysis and calibration error
Antoine Guilcher1, Sally Brett, Shahzad Munir
1Department of Clinical Pharmacology, King's College London British Heart Foundation Centre, St Thomas' Hospital, London, UK.
Objective:
To compare estimation of central cSBP by application of a generalized transfer function (GTF) to a peripheral arterial waveform and from the late systolic shoulder (SBP(2)) of such a waveform and assess errors introduced by noninvasive calibration of the waveform.
Methods:
The digital arterial pulse was acquired noninvasively with a servo-controlled finger cuff. A high fidelity pressure tipped catheter was placed in the proximal aortic root. Measurements were made at baseline (n = 40), after nitrovasodilation, handgrip exercise (n = 18) and during pacing (n = 10). Estimates of cSBP obtained using a GTF and from SBP(2) (using an algorithm applied to individual cardiac cycles) of the digital arterial waveform were compared with values measured at the aortic root.
Results:
When arterial waveforms were calibrated from aortic intra-arterial mean and DBP there was close agreement between estimated and measured cSBP: mean difference between estimated and measured cSBP (SD): 1.0 (5.7) and -0.7 (5.5) mmHg for GTF and SBP(2), respectively. Noninvasive oscillometric calibration increased variability in estimation of cSBP [mean difference, 1.3 (11) mmHg for SBP(2)] but estimates of the cSBP to peripheral systolic pressure increment from oscillometric calibration of SBP(2) agreed well with those obtained using invasive calibration [mean difference -2.4 (6.1) mmHg].
Conclusion:
SBP(2) potentially provides a simple measure of cSBP and is of comparable accuracy to a GTF. Noninvasive calibration increases variability for both methods but has less effect on the cSBP to peripheral SBP increment.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure