Aim:
To asses the image quality and potential for dose reduction of mobile direct detector (DR) chest radiography as compared with computed radiography (CR) for intensive care unit (ICU) chest radiographs (CXR).
Methods And Materials:
Three groups of age-, weight- and disease-matched ICU patients (n=114 patients; 50 CXR per acquisition technique) underwent clinically indicated bedside CXR obtained with either CR (single read-out powder plates) or mobile DR (GOS-TFT detectors) at identical or 50% reduced dose (DR(50%)). Delineation of anatomic structures and devices used for patient monitoring, overall image quality and disease were scored by four readers. In 12 patients pairs of follow-up CR and DR images were available, and in 15 patients pairs of CR and DR(50%) images were available. In these pairs the overall image quality was also compared side-by-side.
Results:
Delineation of anatomy in the mediastinum was scored better with DR or DR(50%) than with CR. Devices used for patient monitoring were seen best with DR, with DR(50%) being superior to CR. In the side-by-side comparison, the overall image quality of DR and DR(50%) was rated better than CR in 96% (46/48) and 87% (52/60), respectively. Inter-observer agreement for the assessment of pathology was fair for CR and DR(50%) (κ = 0.33 and κ = 0.39, respectively) and moderate for DR (κ = 0.48).
Conclusion:
Mobile DR units offer better image quality than CR for bedside chest radiography and allow for 50% dose reduction. Inter-observer agreement increases with image quality and is superior with DR, while DR(50%) and CR are comparable.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...