Development and application of a simple loop-mediated isothermal amplification method on rapid detection of Listeria
1Food Safety Key Laboratory of Guangdong Province, College of Food Science, South China Agricultural University, Guangzhou, 510642, China.
Abstract:
A loop-mediated isothermal amplification (LAMP) method for rapid detection of the food-borne L. monocytogenes strains had been developed and evaluated in this study. The optimal reaction condition was 65 °C for 45 min, with the detection limit as 1 pg DNA/tube and 100 CFU/reaction. Application of the established LAMP assay was performed on 182 food-borne L. monocytogenes strains using a rapid procedure and easy result confirmation, with the sensitivity of LAMP versus PCR assays as 96.7% (176/182) and 91.2% (166/182), respectively; with 100% specificity, positive predictive value (PPV) and negative predictive value (NPV) for both assays.
More Related Videos
08:24Loop-Mediated Isothermal Amplification for Screening Salmonella in Animal Food and Confirming Salmonella from Culture Isolation
Published on: May 20, 2020
07:27The Development of Lyophilized Loop-mediated Isothermal Amplification Reagents for the Detection of Coxiella burnetii
Published on: April 18, 2016
Related Concept Videos
Rapid Identification of Pathogens
Real Time RT-PCR
The real-time quantification of the number of amplified products is...