Related Experiment Video
Updated: Jun 2, 2026

Drug-Induced Sleep Endoscopy (DISE) with Target Controlled Infusion (TCI) and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Comparison of two servo ventilator devices in the treatment of complex sleep apnea
Tomasz J Kuzniar1, Smita Patel, Cynthia L Nierodzik
1Division of Pulmonary and Critical Care Medicine, NorthShore University HealthSystem, Evanston, IL, USA. tkuzniar@northshore.org
Objective:
Servo ventilation (SV) devices generate positive airway pressure with a variable pressure support that changes in response to a patient's own respiratory output. Two currently available SV devices-VPAP-AdaptSV® and BIPAP-AutoSV®-have been used in treatment of complex sleep apnea (CompSAS), but no side-by-side comparisons are available.
Methods:
Data of 76 consecutive patients with complex sleep apnea, who were prescribed a VPAP-AdaptSV® or BIPAP-AutoSV® in a non-randomized parallel design, were retrospectively analyzed. Patients underwent a diagnostic polysomnogram followed by a continuous positive airway pressure (CPAP) titration and a SV titration study. Objective compliance with the device was assessed at the first visit at 4-6weeks of its use.
Results:
Thirty-five patients received a VPAP-AdaptSV® device, while 41 patients were treated with BIPAP-AutoSV®. Patients treated with BIPAP-AutoSV® had a significantly higher apnea-hypopnea index during their CPAP titration study than patients treated with VPAP-AdaptSV® [49/h (28-60) vs. 35/h (19.5-49.5), median (interquartile range), p<0.001]. On follow-up, 56 patients (73.7%) were using their device. Mean nightly use was 5.0h (2.8-6.4) for VPAP-AdaptSV® group and 6.0h (3.5-7.2) for BIPAP-AutoSV® group (p=0.081); an improvement in Epworth Sleepiness Scale score was higher in the BIPAP-AutoSV® group than in the VPAP-AdaptSV® group [4 (1-9) vs. 2.5 (0-5), p=0.02].
Conclusion:
Our retrospective data indicate that the two servo-ventilation devices are comparable means of controlling complex sleep apnea, and the compliance with them is high.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...