Intermittent, erratic behaviour of an implantable cardioverter defibrillator secondary to a hidden magnetic source of
Roy Beinart1, Mary L Guy, Patrick T Ellinor
1Cardiac Arrhythmia Service, Massachusetts General Hospital, 55 Fruit Street, GRB 109, Boston, MA 02114, USA. rbeinart@partners.org
Abstract:
Interference between pacemakers or implantable cardioverter-defibrillators (ICDs) and electromagnetic fields are of great concern due to the potential harmful influence on the function of the device. We report a case of intermittent, erratic behaviour of an ICD secondary to electromagnetic interference (EMI) caused by small magnets that were incorporated into the patient's clothing.