Chest radiography is a poor predictor of left ventricular lead position in patients undergoing cardiac
John Rickard1, Christopher Ingelmo, Dan Sraow
1Tomsich Department of Cardiovascular Medicine, Cleveland Clinic, Heart and Vascular Institute, J1-5, 9500 Euclid Avenue, Cleveland, OH 44195, USA.
Background:
Optimal left ventricular (LV) lead position is thought to be a major predictor of response in patients undergoing cardiac resynchronization therapy (CRT). While the post-implant posterior-anterior (PA) and lateral chest X-ray (CXR) is commonly used to determine the position of the LV lead, the accuracy to which the CXR can correctly localize the LV lead is unknown.
Methods:
We collected data on 47 consecutive patients (mean age 64 years, 60% men and LV ejection fraction 23%, 49% ischemic cardiomyopathy) that underwent CRT between 2004 and 2007, who had both a post-implant CXR as well as a contrast-enhanced multi-detector computed tomography (MDCT) of the chest for any reason. The positions of the LV lead on CXR and MDCT were interpreted in a blinded fashion, independent of each other. The accuracy of the CXR in localizing various LV lead positions, with MDCT as the gold standard, was recorded.
Results:
On CXR, the LV lead tip position was as follows: basal (4%), mid-ventricular (66%), and apical (30%) and anterior (2%), lateral (34%), and posterior (64%). On MDCT, the LV tip position was: basal (28%), mid-ventricular (60%), and apical (13%) and anterior (13%), lateral (19%), and posterior (68%). Compared to the MDCT gold standard, the percentage of LV lead positions the CXR correctly classified correctly were: 100% basal, 39% mid-ventricular, and 29% apical and 0% anterior, 12% lateral, and 77% posterior. Taking both PA and lateral views into consideration, the LV lead position was misclassified by CXR in 62% cases.
Conclusion:
Using MDCT as a gold standard, the routine post-implant CXR performs very modestly in terms of accurate LV lead positioning.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization III: Left Heart Catheterization