Related Experiment Video
Updated: Jun 1, 2026

Making, Testing, and Using Potassium Ion Selective Microelectrodes in Tissue Slices of Adult Brain
Published on: May 7, 2018
Ion-selective optodes measure extracellular potassium flux in excitable cells
Daniel I Harjes1, J Matthew Dubach, Anthony Rosenzweig
1Biomedical Engineering Group, The Charles Stark Draper Laboratory, 555 Technology Square, Cambridge, MA, USA.
Abstract:
Optodes have been used for detection of ionic concentrations and fluxes for several years. However, their uses in biomedical applications have not yet been fully explored. This study investigates optodes as a potential sensor platform for monitoring cellular ion flux with attendant implications in the field of drug screening and toxicology. A prototype system was developed to quantitatively measure extracellular potassium flux from a monolayer of cardiomyocytes. Optodes were created and immobilized on a glass coverslip for fluorescent imaging. The system detected potassium (K(+) ) ion flux during the repolarization phase of the cardiac action potential and further detected a decrease in the magnitude of the flux in the presence of a known K(+) channel inhibitor by optically monitoring local K(+) ion concentrations during field stimulation of the cardiomyocyte monolayer.