A novel RP-HPLC method for simultaneous determination of potassium sorbate and sodium benzoate in soft drinks using
Nafiz O Can1, Goksel Arli, Yigit Lafci
1Department of Analytical Chemistry, Faculty of Pharmacy, Anadolu University, Yunusemre Campus, Eskisehir, Turkey. nafizoc@anadolu.edu.tr.
Abstract:
Potassium sorbate and sodium benzoate are food additives that are generally employed for prevention of food spoilage originating from bacteria, molds or yeasts. Although these compounds were generally recognized as safe due to their low risk of acute and chronic toxicity, they have limitations of usage to protect human health. Development and validation of a novel RP-HPLC method, in which a C18-bonded monolithic silica column was used as stationary phase to assay these compounds, is described for the first time. Aliquots of 10 μL of samples were injected into chromatograph and eluted using phosphate buffer (0.025 M, pH 2.0)-water-acetonitrile (50:45:5, v/v/v) solution, which was pumped at the rate of 3.0 mL/min. To sharpen the peaks, 10 mM octylamine was added to the mobile phase. Potassium sorbate and sodium benzoate were detected at about 12(th) and 14(th) min, respectively, and quantified at 230 nm using photodiode array detector. A total of 41 samples were prepared by simply filtering through 0.45 μm filters after sonication, and injected into the system without any pre-treatment steps. Applicability of the method was demonstrated by performing total procedure on samples of different brands and types, and their compliance to official regulations was assessed.
More Related Videos
11:59Extraction and Detection of Geosmin and 2-Methylisoborneol in Water and Fish using High-Capacity Sorptive Extraction Probes and GC-MS
Published on: July 3, 2025
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Related Concept Videos
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
High-Performance Liquid Chromatography: Elution Process
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...