Bench evaluation of 7 home-care ventilators
Thomas C Blakeman1, Dario Rodriquez, Dennis Hanseman
1Department of Surgery, University of Cincinnati Medical Center, Cincinnati, OH, USA. thomas.blakeman@uc.edu
Background:
Portable ventilators continue to decrease in size while increasing in performance. We bench-tested the triggering, battery duration, and tidal volume (V(T)) of 7 portable ventilators: LTV 1000, LTV 1200, Puritan Bennett 540, Trilogy, Vela, iVent 101, and HT50.
Methods:
We tested triggering with a modified dual-chamber test lung to simulate spontaneous breathing with weak, normal, and strong inspiratory effort. We measured battery duration by fully charging the battery and operating the ventilator with a V(T) of 500 mL, a respiratory rate of 20 breaths/min, and PEEP of 5 cm H(2)O until breath-delivery ceased. We tested V(T) accuracy with pediatric ventilation scenarios (V(T) 50 mL or 100 mL, respiratory rate 50 breaths/min, inspiratory time 0.3 s, and PEEP 5 cm H(2)O) and an adult ventilation scenario (V(T) 400 mL, respiratory rate 30 breaths/min, inspiratory time 0.5 s, and PEEP 5 cm H(2)O). We measured and analyzed airway pressure, volume, and flow signals.
Results:
At the adult settings the measured V(T) range was 362-426 mL. On the pediatric settings the measured V(T) range was 51-182 mL at the set V(T) of 50 mL, and 90-141 mL at the set V(T) of 100 mL. The V(T) delivered by the Vela at both the 50 mL and 100 mL, and by the HT50 at 100 mL, did not meet the American Society for Testing and Materials standard for V(T) accuracy. Triggering response and battery duration ranged widely among the tested ventilators.
Conclusions:
There was wide variability in battery duration and triggering sensitivity. Five of the ventilators performed adequately in V(T) delivery across several settings. The combination of high respiratory rate and low V(T) presented problems for 2 of the ventilators.
More Related Videos
09:37Evaluation of Respiratory Muscle Activation Using Respiratory Motor Control Assessment (RMCA) in Individuals with Chronic Spinal Cord Injury
Published on: July 19, 2013
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome (ARDS)
Published on: April 7, 2021
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...