Optimal slice thickness for cone-beam CT with on-board imager
Kyt Seet1, A Barghi, S Yartsev
1London Regional Cancer Program, London Health Sciences Centre, London, Ontario, Canada.
Purpose:
To find the optimal slice thickness (Δτ) setting for patient registration with kilovoltage cone-beam CT (kVCBCT) on the Varian On Board Imager (OBI) system by investigating the relationship of slice thickness to automatic registration accuracy and contrast-to-noise ratio.
Materials And Method:
Automatic registration was performed on kVCBCT studies of the head and pelvis of a RANDO anthropomorphic phantom. Images were reconstructed with 1.0 ≤ Δτ (mm) ≤ 5.0 at 1.0 mm increments. The phantoms were offset by a known amount, and the suggested shifts were compared to the known shifts by calculating the residual error. A uniform cylindrical phantom with cylindrical inserts of various known CT numbers was scanned with kVCBCT at 1.0 ≤ Δτ (mm) ≤ 5.0 at increments of 0.5 mm. The contrast-to-noise ratios for the inserts were measured at each Δτ.
Results:
For the planning CT slice thickness used in this study, there was no significant difference in residual error below a threshold equal to the planning CT slice thickness. For Δτ > 3.0 mm, residual error increased for both the head and pelvis phantom studies. The contrast-to-noise ratio is proportional to slice thickness until Δτ = 2.5 mm. Beyond this point, the contrast-to-noise ratio was not affected by Δτ.
Conclusion:
Automatic registration accuracy is greatest when 1.0 ≤ Δτ (mm) ≤ 3.0 is used. Contrast-to-noise ratio is optimal for the 2.5 ≤ Δτ (mm) ≤ 5.0 range. Therefore 2.5 ≤ Δτ (mm) ≤ 3.0 is recommended for kVCBCT patient registration where the planning CT is 3.0 mm.
More Related Videos
07:28Thinned-skull Cortical Window Technique for In Vivo Optical Coherence Tomography Imaging
Published on: November 19, 2012
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...