Optimal slice thickness for cone-beam CT with on-board imager
Kyt Seet1, A Barghi, S Yartsev
1London Regional Cancer Program, London Health Sciences Centre, London, Ontario, Canada.
Optimal kilovoltage cone-beam CT (kV-CBCT) slice thickness for patient registration is 2.5-3.0 mm. This range balances accurate registration with good contrast-to-noise ratio for Varian OBI systems.
Area of Science:
- Medical Physics
- Radiological Imaging
- Radiation Oncology
Background:
- Accurate patient positioning is critical for effective radiation therapy.
- Kilovoltage cone-beam CT (kV-CBCT) is widely used for image-guided radiation therapy (IGRT).
- Slice thickness impacts kV-CBCT image quality and registration accuracy.
Purpose of the Study:
- To determine the optimal slice thickness (Δτ) for kV-CBCT patient registration on a Varian OBI system.
- To investigate the relationship between slice thickness, automatic registration accuracy, and contrast-to-noise ratio (CNR).
Main Methods:
- Automatic registration accuracy was assessed using head and pelvis phantom studies with known offsets.
- Images were reconstructed with slice thicknesses ranging from 1.0 to 5.0 mm.
- Contrast-to-noise ratio was measured using inserts with known CT numbers at various slice thicknesses.
Main Results:
- Registration residual error increased for slice thicknesses greater than 3.0 mm.
- Contrast-to-noise ratio was proportional to slice thickness up to 2.5 mm.
- No significant difference in residual error was observed below the planning CT slice thickness threshold.
Conclusions:
- Automatic registration accuracy is maximized with slice thicknesses between 1.0 mm and 3.0 mm.
- Optimal contrast-to-noise ratio is achieved with slice thicknesses between 2.5 mm and 5.0 mm.
- A slice thickness of 2.5-3.0 mm is recommended for kV-CBCT patient registration when using 3.0 mm planning CT.
More Related Videos
07:28Thinned-skull Cortical Window Technique for In Vivo Optical Coherence Tomography Imaging
Published on: November 19, 2012
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...