Appearance of discontinuities in spectra transformed by the piecewise direct instrument standardization procedure
P J Gemperline1, J Cho, P K Aldridge
1Department of Chemistry, East Carolina University, Greenville, North Carolina 27858.
Abstract:
Several years ago, we noted that spectra transformed by the piecewise direct standardization (PDS) method may contain discontinuities. Having noticed that the problem was a recurring one, we studied it and recently diagnosed its source. Our investigations suggest that this problem also occurs in applications of window factor analysis, evolving factor analysis, and any other procedure that uses piecewise principal component models. In this work, we report the source of the problem and illustrate it with one example. A procedure is presented for eliminating the problem that is effective in PDS pattern recognition applications. Further work is needed to develop modified algorithms suitable for calibration applications.
More Related Videos
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Discrete Fourier Transform
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.