Simultaneous grayscale and subharmonic ultrasound imaging on a modified commercial scanner
J R Eisenbrey1, J K Dave, V G Halldorsdottir
1Department of Radiology, Thomas Jefferson University, Philadelphia, PA 19107, United States.
Objective:
To demonstrate the feasibility of simultaneous dual fundamental grayscale and subharmonic imaging on a modified commercial scanner.
Motivation:
The ability to generate signals at half the insonation frequency is exclusive to ultrasound contrast agents (UCA). Thus, subharmonic imaging (SHI; transmitting at f(0) and receiving at f(0)/2) provides improved visualization of UCA within the vasculature via suppression of the surrounding tissue echoes. While this capability has proven useful in a variety of clinical applications, the SHI suppression of surrounding tissue landmarks (which are needed for sonographic navigation) also limits it use as a primary imaging modality. In this paper we present results using a commercial ultrasound scanner modified to allow imaging in both grayscale (f(0)=4.0 MHz) and SHI (f(0)=2.5 MHz, f(0)/2=1.25 MHz) modes in real time.
Methods:
A Logiq 9 ultrasound scanner (GE Healthcare, Milwaukee, WI) with a 4C curvilinear probe was modified to provide this capability. Four commercially available UCA (Definity, Lantheus Medical Imaging, North Billerica, MA; Optison, GE Healthcare, Princeton, NJ; SonoVue, Bracco Imaging, Milan, Italy; and Sonazoid, GE Healthcare, Oslo, Norway) were all investigated in vitro over an acoustic output range of 3.34 MPa. In vivo the subharmonic response of Sonazoid was investigated in the portal veins of four canines (open abdominal cavity) and four patients with suspected portal hypertension.
Results:
In vitro, the four UCA showed an average maximum subharmonic amplitude of 44.1±5.4 dB above the noise floor with a maximum subharmonic amplitude of 48.6±1.6 dB provided by Sonazoid. The average in vivo maximum signal above the noise floor from Sonazoid was 20.8±2.3 dB in canines and 33.9±5.2 dB in humans. Subharmonic amplitude as a function of acoustic output in both groups matched the S-curve behavior of the agent observed in vitro. The dual grayscale imaging provided easier sonographic navigation, while the degree of tissue suppression in SHI mode varied greatly on a case by case basis.
Conclusions:
These results demonstrate the feasibility of dual grayscale and SHI on a modified commercial scanner. The ability to simultaneously visualize both imaging modes in real time should improve the applicability of SHI as a future primary clinical imaging modality.
More Related Videos
12:00Non-invasive Parenchymal, Vascular and Metabolic High-frequency Ultrasound and Photoacoustic Rat Deep Brain Imaging
Published on: March 2, 2015
06:02Multi-timescale Microscopy Methods for the Characterization of Fluorescently-labeled Microbubbles for Ultrasound-Triggered Drug Release
Published on: June 12, 2021
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Imaging Studies II: Ultrasonography
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...