Trabecular Bone Microarchitecture Evaluation in an Osteoporosis Mouse Model
Published on: September 8, 2023
MicroCT morphometry analysis of mouse cancellous bone: intra- and inter-system reproducibility
K Verdelis1, L Lukashova, E Atti
1Mineralized Tissues Laboratory, The Hospital for Special Surgery, New York, NY, USA. kv100@pitt.edu
Bone
|May 31, 2011
Summary
Microcomputed tomography (microCT) systems show variable performance in mouse bone analysis. Direct comparison of trabecular morphometry measurements across different microCT systems is not recommended due to significant differences.
Area of Science:
- Biomedical Engineering
- Materials Science
- Orthopedics
Background:
- Microcomputed tomography (microCT) is crucial for 3D bone analysis.
- Reproducibility and agreement of microCT measurements across systems, especially at high resolutions, are not well-established.
Purpose of the Study:
- To compare the performance reproducibility of three ex vivo microCT systems.
- To evaluate morphometric measurements of mouse cancellous bone at different voxel sizes.
Main Methods:
- 3D morphometric analyses of mouse distal femoral epiphyses were performed using GE eXplore Locus SP, Scanco μCT35, and Skyscan 1172 microCT systems.
- Scans were conducted in triplicate at 12 μm and 8 μm voxel sizes.
- Trabecular bone morphometry parameters were analyzed, examining relative values, voxel size dependence, and intra-system reproducibility.
Main Results:
- Bone volume fraction and trabecular separation showed agreement across systems.
- Relative bone surface, trabecular thickness, and number varied significantly among the three microCT systems.
- Trabecular thickness and number showed the largest discrepancies (up to 236% and 218%, respectively).
- Voxel size dependence varied, particularly for trabecular number (0-20% difference).
- Intra-system reproducibility was variable and generally improved with smaller voxel sizes for some parameters and systems.
Conclusions:
- Direct comparison of trabecular morphometry measurements obtained from different microCT systems is not advisable.
- Optimal microCT system and voxel size selection for mouse bone studies should be based on specific measurement interests and the system used.