Performance of the PT100 noncontact tonometer in healthy eyes
1Cornea Research Chair, Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh, Kingdom of Saudi Arabia. turkim@ksu.edu.sa
Background:
The purpose of this study was to assess the repeatability and reproducibility of the PT100 noncontact tonometer and to compare its consistency with the Goldmann applanation tonometer (GAT) in measuring intraocular pressure (IOP).
Methods:
Triplicate IOP measurements were obtained on two separate occasions using the PT100 and GAT from randomly selected eyes in 66 healthy volunteers aged 22 ± 1 years. The repeatability and reproducibility of each techniques was assessed. Agreement between the techniques was statistically quantified using intrasession repeatability for each technique as the basis for comparison.
Results:
Both techniques returned equal IOP values in the first measurement session (15 ± 3 mmHg). The second session showed a mean difference in average IOP (1 ± 0.71). The 95% limits of agreement between the techniques were -5.2 to 5.5 mmHg and -4.0 to 4.7 mmHg (sessions 1 and 2, respectively). These mean differences were not statistically significant (P > 0.05, paired t-test), with the PT100 underestimating IOP measurement by 1.00 mmHg. The mean intrasession IOP for GAT sessions 1 and 2 was 0 ± 0.90 mmHg and 0.04 ± 1.06 mmHg, respectively, and the corresponding mean IOP measurement difference for the PT100 was -0.06 ± 0.96 and -0.39 ± 0.94 mmHg (sessions 1 and 2, respectively; P > 0.05, paired t-test). Repeatability coefficients for the GAT IOP measurements were 1.8 mmHg and 2.1 mmHg for sessions 1 and 2, while the PT100 repeatability coefficient was 1.9 mmHg and 1.8 mmHg for sessions 1 and 2, respectively. The intrasession repeatability coefficient of both techniques for test-retest differences were within ±5 mmHg.
Conclusion:
The PT100 noncontact tonometer produced greater repeatability than the GAT in assessment of IOP, whereas GAT resulted in more reproducible results. Both techniques showed a close level of agreement on comparison, with the PT100 underestimating IOP measurement by 1.0 mmHg only, although this was not clinically or statistically significant. Of importance is that the IOP measurements using these techniques could be interchangeable in the IOP range studied here.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...