On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
Optimization of Chemiluminescent Nitrogen Detection for Packed-Column Supercritical Fluid Chromatography with
M T Combs1, M Ashraf-Khorassani, L T Taylor
1Department of Chemistry, Virginia Polytechnic Institute and State University, Blacksburg, Virginia 24061.
Abstract:
A chemiluminescent nitrogen detector was optimized for packed-column supercritical fluid chromatography. Methanol modifier concentrations of 15% or less, an ozone flow of 5.8 mL/min, and a decompressed CO(2) flow between 240 and 310 mL/min were found to exhibit a maximum sensitivity of 5 ng for sulfamethazine (1 ng of N). The addition of a membrane drier to the "optimized" system further decreased the minimum detectable quantity (MDQ) to 0.5 ng (0.1 ng of N). In addition, by using a microbore column (2 mm i.d.) instead of an analytical scale column, the postcolumn decompressed flow split could be eliminated, further reducing the MDQ to 0.125 ng of sulfamethazine (0.025 ng of N).
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Instrumentation