X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Effect of contrast material on image noise and radiation dose in adult chest computed tomography using automatic
Jijo Paul1, Boris Schell, J Matthias Kerl
1Clinic of the Goethe University, Department of Diagnostic and Interventional Radiology, Haus 23C UG, Theodor-Stern-Kai 7, 60590 Frankfurt am Main, Germany. jijopaul1980@gmail.com
Purpose:
To determine the difference in radiation dose between non-enhanced (NECT) and contrast-enhanced (CECT) chest CT examinations contributed by contrast material with different scanner generations with automatic exposure control (AEC).
Methods & Materials:
Each 42 adult patients received a NECT and CECT of the chest in one session on a 16-, 64- or 128-slice CT scanner with the same scan protocol settings. However, AEC technology (Care Dose 4D, Siemens) underwent upgrades in each of the three scanner generations. DLP, CTDIvol and image noise were compared.
Results:
Although absolute differences in image noise were very small and ranged between 10 and 13 HU for NECT and CECT in median, the differences in image noise and dose (DLP: 16-slice:+2.8%; 64-slice:+3.9%; 128-slice:+5.6%) between NECT and CECT were statistically significant in all groups. Image noise and dose parameters were significantly lower in the most recent 128-slice CT generation for both NECT and CECT (DLP: 16-slice:+35.5-39.2%; 64-slice:+6.8-8.5%).
Conclusion:
The presence of contrast material lead to an increase in dose for chest examinations in three CT generations with AEC. Although image noise values were significantly higher for CECT, the absolute differences were in a range of 3 HU. This can be regarded as negligible, thus indicating that AEC is able to fulfill its purpose of maintaining image quality. However, technological developments lead to a significant reduction of dose and image noise with the latest CT generation.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging