False EX1 signatures caused by sample carryover during HX MS analyses
Jing Fang1, Kasper D Rand, Penny J Beuning
1Department of Chemistry & Chemical Biology, Northeastern University, Boston, MA 02115.
False EX1 kinetics in hydrogen exchange mass spectrometry (HX-MS) can be caused by peptide carryover. This study identifies carryover origins, explains its impact on HX-MS data, and presents an optimized washing protocol to eliminate it for reliable deuterium incorporation analysis.
Area of Science:
- Analytical Chemistry
- Biochemistry
- Structural Biology
Background:
- Hydrogen exchange mass spectrometry (HX-MS) is a powerful technique for studying protein dynamics and structure.
- The EX1 kinetic model describes rapid deuterium incorporation upon simultaneous unfolding and solvent exposure of multiple residues.
- False EX1 kinetics can arise from experimental artifacts, complicating data interpretation.
Purpose of the Study:
- To investigate the phenomenon of false EX1 kinetics in HX-MS experiments.
- To identify peptide carryover as a significant contributor to false EX1.
- To develop and validate a method for eliminating peptide carryover in HX-MS.
Main Methods:
- Exploration of peptide carryover origins in chromatographic separations used in HX-MS.
- Analysis of mass spectra to understand how peptide carryover mimics EX1 kinetics.
- Development and testing of an optimized solvent washing protocol to remove residual peptides.
Main Results:
- Peptide carryover between chromatographic runs was identified as a primary cause of false EX1 signatures in HX-MS.
- A systematic series of solvent injections was optimized to effectively remove peptide carryover.
- The optimized washing protocol successfully eliminated carryover artifacts, enabling reliable deuterium incorporation measurements for proteins exhibiting high carryover.
Conclusions:
- Peptide carryover is a critical artifact that must be addressed for accurate HX-MS data interpretation.
- The developed washing protocol provides an effective solution for mitigating peptide carryover in HX-MS.
- This methodology ensures the reliability of deuterium incorporation studies, particularly for challenging protein samples.
More Related Videos
08:17Practical Aspects of Sample Preparation and Setup of 1H R1ρ Relaxation Dispersion Experiments of RNA
Published on: July 9, 2021
08:22A Robust Polymerase Chain Reaction-based Assay for Quantifying Cytosine-guanine-guanine Trinucleotide Repeats in Fragile X Mental Retardation-1 Gene
Published on: September 16, 2019
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Tandem Mass Spectrometry
¹H NMR: Complex Splitting
Splitting diagrams or splitting tree diagrams are routinely used to depict such complex couplings. While drawing splitting diagrams, the splitting with the larger coupling constant is usually applied first.
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
¹H NMR Signal Multiplicity: Splitting Patterns
Interpreting ¹H NMR Signal Splitting: The (n + 1) Rule